PC Digitizer Performance Verification
To verify the PC digitizer card input's performance (continued) (continued)
Step
2 Open the E5505A phase noise
E5500_user_interface
04 Apr 04 rev 1
3 Click FFT Analyzer Check I/O button
E5500_server_hardware3
04 Apr 04 rev 1
4 Configure the function generator
450
software.
in Server Hardware Connections to
verify the PC digitizer's connectivity.
using the appropriate keys on the
instrument.
Action
Path: Start\Programs\Agilent Subsystems\E5500
Phase Noise System\E5500 User Interface
Path: System\Server Hardware Connections
A green check-mark appears on the button to
confirm connectivity. If a red circle with a slash
appears, check hardware connections and try again.
Frequency: 100 kHz
Amplitude: 100 mV peak to peak
For instructions, refer to the function generator's
user's guide.
Agilent E5505A User's Guide