Figure 105 Connect diagram for the free-running RF oscillator measurement
4
The test set's signal input is subject to the limits and characteristics in
C A UTI ON
page 153.
To prevent damage to the test set's components, do not apply the input signal to the
test set's signal input connector until the input attenuator (Option 001) has been set
by the phase noise software, which will occur when the connection diagram
appears.
152
When the Connect Diagram dialog box appears, click on the hardware
drop-down arrow and select your hardware configuration from the list. See
Figure
105.
TEST SET
N5500A
Connect your DUT and reference sources to the test set at this time.
Confirm your connections as shown in the connect diagram.
The input attenuator (Option 001 only) is now correctly configured
based on your measurement definition.
DOWNCONVERTER
N5502A
on
Table 18
Agilent E5505A User's Guide