Figure 274. Measured Loop Suppression Curve - Agilent Technologies e1420b User Manual

Phase noise measurement system
Hide thumbs Also See for e1420b:
Table of Contents

Advertisement

15
Advanced Software Features
There are four different curves available for this graph:
.
e5505a_user_loop_suppression.ai
rev2 10/24/03
Figure 274 Measured loop suppression curve
348
"Measured" loop suppression curve
a
result of the loop suppression measurement performed by the E5505A
system.
"Smoothed" measured suppression curve
b
is a curve-fit representation of the measured results, it is used to
compare with the "theoretical" loop suppression.
c
"Theoretical" suppression curve
predicted loop suppression based on the initial loop parameters
defined/selected for this particular measurement (kphi, kvco, loop
bandwidth, filters, gain, etc.).
"Adjusted" theoretical suppression curve
d
through
Figure 279
value of suppression for this measurement. It is based on changing loop
parameters (in the theoretical response) to match the "smoothed"
measured curve as closely as possible.
(Figure 274
(Figure 276
on page 351) —this is the new "adjusted" theoretical
on page 348)—this is the
(Figure 275
on page 349)—this
on page 349)—this is the
(Figure 277
on page 350