Phase noise measurement system
Hide thumbs Also See for e1420b:
Table of Contents

Advertisement

If the output amplitude of your DUT is not sufficient to provide an adequate
measurement noise floor, it will be necessary to insert a low-noise amplifier
between the DUT and the test set. Refer to "Inserting an Device" in
"Absolute Measurement Fundamentals for details on determining the effect
the amplifiers noise will have on the measured noise floor.
VCO reference
In order for the noise measurement results to accurately represent the noise of
the DUT, the noise level of the reference source should be below the expected
noise level of the DUT.

Beginning the measurement

1
.
E5500_new_measurement
04 Apr 04 rev 1
Figure 103 Selecting a new measurement
2
E5500_new_cali_meas
04 Apr 04 rev 1
Figure 104 Confirm measurement dialog box
Agilent E5505A User's Guide
From the Measurement menu, choose New Measurement. See
When the Do you want to perform a New Calibration and Measurement?
prompt appears, click Yes.
Absolute Measurement Examples
Chapter
Figure
6
5,
103.
151