Rf Synthesizer Using Efc; Required Equipment; Defining The Measurement - Agilent Technologies e1420b User Manual

Phase noise measurement system
Hide thumbs Also See for e1420b:
Table of Contents

Advertisement

RF Synthesizer Using EFC

This measurement example will help you measure the absolute phase noise of
an RF synthesizer using EFC.
To prevent damage to the test set's components, the input signal do not apply the
C A UTI ON
signal input connector until the input attenuator has been correctly set for the
desired configuration, as shown in Table 31. Apply the input signal when the
connection diagram appears

Required equipment

This measurement requires an Agilent 8257x with an EFC Input port, in
addition to the phase noise test system and your DUT. (For more information,
refer to the section
the coaxial cables and adapters necessary to connect the DUT and reference
source to the test set.
To ensure accurate measurements allow the DUT and measurement equipment to warm up
NO T E
at least 30 minutes before making the noise measurement

Defining the measurement

1
2
3
e5505a_user_select_param_def_EFCprn
24 Jun 04 rev 3
Figure 121 Select the parameters definition file
Agilent E5505A User's Guide
"Selecting a reference source"
From the File menu, choose Open.
If necessary, choose the drive or directory where the file you want is stored.
In the File Name box, choose "RFSynth_EFC.pnm". See
Absolute Measurement Examples
on page 132.) You also need
Figure
121.
6
173