Figure 44 Connection diagram
4
The test set's signal input is subject to the limits and characteristics contained in
C A UTI ON
Table 6
To prevent damage to the test set's hardware components, do not apply the input
signal to the test set's signal input connector until the input attenuator (Option 001)
has been set by the phase noise software, which occurs at the connection diagram.
5
6
72
Connect your DUT and reference sources to the test set at this time and
confirm your connections as shown in the appropriate connect diagram.
The input attenuator (Option 001 only) is now correctly configured
based on your measurement definition.
on page 73.
Press Continue.
As the system performs the calibration routines, various status messages
appear on the display. When the last message—Measuring PLL
suppression—appears, you can choose to continue the routine or stop it.
Press Continue using Adjusted Loop Suppression to continue making the
noise measurement, or press Abort to stop the measurement. (For
descriptions of the messages, see
time to read through the descriptions while the system completes the
routines.)
"Status messages"
on page 74. You have
Agilent E5505A User's Guide