17
System Specifications

Specifications

This section contains mechanical and environmental specifications, operating
characteristics, power requirements, and PC requirements for the system. It
also provides specifications for accuracy, measurement qualifications, and
tuning.
Table 66
system.
Table 66 Mechanical and environmental specifications
Specifications
Temperature:
Operating
Nonoperating/storage
Altitude
Maximum relative humidity
Air flow space required
System weight:
Approximate, typical
System dimensions:
Height
Width
with rack-mounted flat-panel display
with benchtop PC and display
Depth
Table 67 Operating characteristics
Warm up time required
Carrier frequency ranges
Offset frequency ranges
System noise response
System spurious response
388
contains the mechanical and environmental specifications for a
Table 67
shows the system's operating characteristics.
Values
05 °C to 40 °C (41 °F to 104 °F)
-40 °C to 70 °C (-40 °F to 158 °F)
Up to 2,000 m (6,500 ft)
80% for temperatures up to 31 °C, decreasing linearly to 50% relative
humidity at 40 °C
102 mm (4 in) on all sides
Benchtop
210 lbs (95.5 kg)
711 to 915 mm (28 to 36 in)
432 mm (17 in)
n/a
1,016 mm (40 in)
711 mm (28 in)
1.6 Meter Rack
2 Meter Rack
431 lbs (195.5 kg)
465 lbs (211 kg)
1,620 mm (64 in)
2,020 mm (79.5 in)
600 mm (23.6 in)
600 mm (23.6 in)
1438 mm (56.7 in)
1438 mm (56.7 in)
n/a
n/a
905 mm (35.7 in)
905 mm (35.7 in)
30 minutes
50 KHz to 1.6 GHz
50 KHz to 6 GHz
50 KHz to 18 GHz
50 KHz to 26.5 GHz
0.01 Hz to 2 MHz
0.01 Hz to 100 MHz
–180 dBc/Hz typically (>10 kHz offsets)
≤120 dBc typically