Figure 223. Am Noise Calibration Setup - Agilent Technologies e1420b User Manual

Phase noise measurement system
Hide thumbs Also See for e1420b:
Table of Contents

Advertisement

E5505a_am_noise_cal_setup
01 Mar 04 rev 1
Figure 223 AM noise calibration setup
3
4
The quadrature meter should be at zero volts due to the blocking capacitor at the AM
NO T E
detector's output.
Figure 224 AM detector sensitivity graph
Agilent E5505A User's Guide
Power meter
or spectrum
analyzer
DUT
Locate the drive level on the AM sensitivity graph
the data.
Measure the noise data and interpret the results. The measured data will be
plotted as single-sideband AM noise in dBc/Hz.
e5505a_user_AM_cal.ai
rev2 10/24/03
AM Noise Measurement Fundamentals
(Figure
224), and enter
11
291