If the output amplitude of your DUT is not sufficient to provide an adequate
measurement noise floor, it is necessary to insert a low-noise amplifier
between the DUT and the test set. Refer to
details on determining the effect the amplifiers noise will have on the
measured noise floor.

Beginning the measurement

1
E5500_new_measurement
04 Apr 04 rev 1
Figure 42 Selecting new measurement
2
E5500_new_cali_meas
Figure 43 Confirm new measurement
3
Agilent E5505A User's Guide
From the Measurement menu, choose New Measurement. See
When the Do you want to perform a New Calibration and Measurement?
prompt appears, click Yes.
When the Connect Diagram dialog box appears, click on the hardware
pull-down arrow and select your hardware configuration from the list. See
Figure 44
on page 72.
Expanding Your Measurement Experience
"Inserting a Device"
4
on page 122 for
Figure
42.
71