Residual Measurement Examples
The test set's signal input is subject to the limits and characteristics contained in
C A UTI ON
Table
To prevent damage to the test set's hardware components, do not apply the input
signal to the test set's signal input connector until the input attenuator (Option 001)
has been set by the phase noise software, which occurs at the connection diagram.
Table 36 Test set signal input limits and characteristics
Limits
Frequency
Maximum Signal Input Power
At Attenuator Output, Operating Level
Range:
RF Phase Detectors
Microwave Phase Detectors
Characteristics
Input Impedance
Refer to the following system connect diagram examples in
NO T E
Interconnections" for more information about system interconnections.
Figure
Figure
Figure
236
36.
307, "E5505A system connections with standard test set," on page 401
308, "E5505A system connections with test set option 001," on page 402
309, "E5505A system connections with test set option 201," on page 403
50 kHz to 1.6 GHz (Std)
50 kHz to 26.5 GHz (Option 001)
50 kHz to 26.5 GHz (Option 201)
Sum of the reference and signal input power shall not
exceed +23 dBm (+30 dBm for
Option 001)
0 to +23 dBm (Signal Input)
+15 to +23 dBm (Reference Input)
0 to +5 dBm (Signal Input)
+7 to +10 dBm (Reference Input)
50 Ω nominal
Chapter
18, "System
Agilent E5505A User's Guide

Advertisement

Table of Contents