RF Interface Cable Connections

Figure 10
how it should be configured to the E8364C 2-Port and N5230C PNA Network Analyzer.
1. Remove the SOURCE OUT to CPLR THRU and RCVR IN to CPLR ARM jumpers (x8)
on the PNA. The RCVR R1 - R2 to SOURCE OUT reference loop jumpers (x2) remain on
the front panel.
2. Connect the RF cables supplied with this option between the network analyzer and the
U3025AE10 Test Set. Torque each cable to 8 in-lb. Refer to
page
CAUTION
CAUTION
Table 5
RF Cables
U3025-20044
U3025-20044
U3025-20045
U3025-20045
Figure 10
User's Guide U3025-90001
This Manual:http://www.manuallib.com/agilent/u3025ae10-test-set-user-guide.html
illustrates the setup configuration of the U3025AE10 Multiport Test Set and
15.
Each end of the interconnect RF cables have a different length from the bend.
When connecting the RF Interconnect cables be sure that the longer end from
the bend is connected to the PNA.
Over torque will cause damage to the Test Set and may cause connectors to
spin or become loose.
12-Port Cable Connection
From (PNA)
SOURCE OUT
CPLR THRU
CPLR ARM
RCVR A IN, B IN
12-Port Setup Configuration
System Setup E8363C, E8364C or N5230C
To (Test Set)
SOURCE OUT
CPLR THRU
CPLR ARM
A IN, B IN
Table 5
and
Figure 10 on
45 (x4)
44 (x4)
44
45