Parameter Overview 7 - ABB Advance Optima magnos206 Operator's Manual

Ao2000 series gas analyzers
Table of Contents

Advertisement

Parameter Overview
Figure 7-D-4
Parameter Overview
Explanations
Name
Measuring point
Source
Page
Position
Style
Range Low,
Range High
Places
42/24-10 EN Rev. 9
The display parameters have the following functions:
The Name of the display element is set by the system and cannot be changed.
The description entered for the Measuring point appears over the display
element during measurement operation. The description is set by the system; it
can be changed for the user-configured display elements. The maximum length is
20 characters.
The Source of the display elements is always a function block. The source
cannot be changed for the display elements of the default assignment, i.e. the
measured quantities, and for the key entries.
The parameter Page indicates the page on which the display element is shown.
Each display element can be moved to any system or user page.
The Position of a display element on a system page is determined by the
system. It can be changed by being exchanged with another display element.
The user can freely configure the position on a user page.
The Style of display depends on the source type. The following display styles
exist: bar display, point display, value entry and key entry.
Examples of the different display styles are shown as soon as this parameter is
selected.
The value entry and key entry display styles are explained in greater detail on
pages 7-D-10 and 7-D-12.
The Range Low and Range High parameters determine the measurement
range of the bar display and the point display. They cannot be changed for the
display elements of the default assignment, i.e., the measured quantities.
The Places parameter determines the number of decimal places for the digital
display of the measured values for the bar display and point display. It cannot be
changed for the display elements of the default assignment, i.e., the measured
quantities.
Chapter 7: Gas Analyzer Configuration
7-D-5