4-Way Analog Output Module: Electrical Connections - ABB Advance Optima magnos206 Operator's Manual

Ao2000 series gas analyzers
Table of Contents

Advertisement

4-Way Analog Output Module: Electrical Connections

Figure 4-5
4-Way Analog Output
Module
Connection Diagram
42/24-10 EN Rev. 9
AO1-AO4
Analog Outputs:
0/4-20 mA (configurable, factory-set to 4-20 mA), common negative
pole, galvanically separated from ground, freely connectable to
ground, max. gain vs. local protective ground potential 50 V,
max. working resistance 750 Ω. Resolution 16 bit. The output signal
cannot be lower than 0 mA.
Design: 8-pin terminal strip for stranded or solid cable with a maximum section of
2
1 mm
(17 AWG). Observe the information on needed materials (see page 4-13).
"Standard Terminal Connections" section (see page 4-9) contains information on
standard terminal assignment. Please observe also the specifications in the
analyzer data sheet.
Chapter 4: Electrical Connection
1
2
3
4
5
6
7
0/4 to 20 mA, max. 750 Ω
1
AO1
+
-
2
0/4 to 20 mA, max. 750 Ω
3
AO2
+
4
-
0/4 to 20 mA, max. 750 Ω
5
AO3
+
6
-
0/4 to 20 mA, max. 750 Ω
7
AO4
+
8
-
8
4-5