ABB Advance Optima magnos206 Operator's Manual page 218

Ao2000 series gas analyzers
Table of Contents

Advertisement

Status Messages,
No. Text
336 A computer error occurred
during calculation of the
measurement value.
337 Broken analog output line.
338 Broken digital input line
(moisture sensor).
339 Broken or shorted analog
input line.
340 Analog input value above or
below upper or lower limit
value 1.
341 Analog input value above or
below upper or lower limit
value 2.
Flow Monitor (Gas Module)
342 Flow rate under limit value 1.
343 Flow rate under limit value 2.
Measurement Value
344 Value above measurement
value range.
345 Value below measurement
value range.
Limas
356 Analyzer in warm-up phase.
357 Limas motor optimization in
progress.
358 Lamp intensity above or below
middle of permissible range.
359 Lamp intensity above or below
permissible range.
10-12
continued
O E M F Reaction / Remark
x x
Check the configuration and calibration of the analog
inputs and outputs.
x x
Check the analog output lines.
x x
Check the moisture sensor in the system cooler.
x x
Check system cooler temperature.
x
Check system cooler temperature.
x
Check system cooler temperature.
x
Check sample preparation.
Limit value 1 = 25% MRS.
x x
Check sample preparation.
Limit value 2 = 10 % MRS.
Automatic calibration is interrupted and locked out.
Value > +130 % MRS, not logged
Value < –100 % MRS, not logged
x x
Status message during warm-up phase.
If the status message appears after warm-up, a
temperature error has occurred in the lamp or in the sample
cell or in the measurement or reference detector amplifier.
Notify service.
x x
Status message after warm-up phase
x
Lamp intensity has fallen to 10% of Init value. Check the
intensity values in the Diagnostics/Test
specific
If all four values have fallen by about the same amount
compared to the Init values, the cause is reduced lamp
intensity. The lamp should soon be replaced. If only the
two measurement detector values have dropped, the
cause is probably a contaminated sample cell. Clean the
sample cell or exchange it.
x x
Lamp intensity has fallen to 5% of Init value. Check the
intensity values in the Diagnostics/Test
specific
have fallen by about the same amount compared to the Init
values, the cause is reduced lamp intensity. Replace the
lamp and perform an amplification optimization.
Chapter 10: Status Messages, Troubleshooting
Lamp Intensity menu.
Lamp Intensity menu. If all four values
Module
Module
Continued on next page
42/24-10 EN Rev. 9