ABB Advance Optima magnos206 Operator's Manual page 221

Ao2000 series gas analyzers
Table of Contents

Advertisement

Status Messages,
No. Text
509 Automatic calibration started.
510 Automatic calibration ended.
511 Automatic calibration
externally interrupted.
512 Automatic calibration in
progress.
513 System bus communication
defect during automatic
calibration.
514 External calibration started.
515 External calibration ended.
516 External calibration in
progress.
517 Device being serviced.
518 The calibration could not be
done, because the value is not
stable.
519 Preamplifier overflow error:
Calibration could not be
performed because of
preamplifier override.
520 Initial zero calibration started.
521 Initial zero calibration ended.
522 Initial zero calibration
interrupted.
523 Initial zero calibration
incomplete. System bus
communication defect during
calibration.
524 Initial zero calibration in
progress.
525 Linearization impossible:
Linearization did not produce
a valid result. Measurement
value possibly inaccurate.
Check center point gas.
526 Linearization impossible:
Linearization could not be
performed Ie the characteristic
is linear.
527 Initial calibration for
component:
528 Autocalibration not started Ie
manual calibration was
running.
42/24-10 EN Rev. 9
continued
O E M F Reaction / Remark
For information
For information
For information
x For information, not logged
For information
For information
x For information, not logged
x For information, e.g. during manual calibration, not logged
For information
For information
For information
For information
x For information, not logged
See message text
See message text
For information
For information
Chapter 10: Status Messages, Troubleshooting