ABB Advance Optima magnos206 Operator's Manual page 217

Ao2000 series gas analyzers
Table of Contents

Advertisement

Status Messages,
No. Text
MultiFID
321 The detector temperature is
below the lowest permissible
temperature.
322 The flame is out.
323 The analyzer is in the fail-safe
state.
Temperature Regulator
324 Temperature above or below
upper and/or lower limit
value 1.
325 Temperature above or below
upper and/or lower limit
value 2.
Pressure Regulator
326 No new measurement values
from analog/digital converter.
327 Measurement value exceeds
the analog/digital converter
value range.
328 A computer error occurred
during calculation of the
measurement value.
329 Pressure above or below
upper and/or lower limit
value 1.
330 Pressure above or below
upper and/or lower limit
value 2.
331 The pressure regulator control
variable is beyond the valid
range.
I/O Devices
332 Accessory voltage defect on
I/O board.
333 Unavailable I/O type
configured.
334 No new measurement values
from analog/digital converter.
335 Measurement value exceeds
the analog/digital converter
value range.
42/24-10 EN Rev. 9
continued
O E M F Reaction / Remark
x x
Status message during the warm-up phase.
If the status message appears after warm-up:
Check the thermal link and replace as needed (see the
MultiFID14 analyzer module start-up and maintenance
manual, publication number 41/24-105 EN).
x x
Status message during the warm-up phase.
If the status message appears after warm-up:
Check the gas supply.
x x
Inadequate combustion gas supply; turn power supply off
and back on after ≥ 3 seconds.
Notify service.
x
Status message during the warm-up phase.
If the status message appears after warm-up:
Check if the permissible ambient temperature range is
being maintained (see page 10-19). Check the analyzer
x
module thermal link and replace if necessary.
x x
Notify service.
x x
Notify service.
x x
Notify service.
x
MultiFID14: Check the supply gas pressure.
x
MultiFID14: Check the supply gas pressure.
x x
MultiFID14: Check the supply gas pressure.
x x
Defective I/O board. Replace the board.
x x
Correct the configuration with the test and calibration
software.
x x
Defective I/O board. Replace the board.
x x
Check signals at analog inputs.
If OK, check the configuration and calibration of the analog
inputs.
Chapter 10: Status Messages, Troubleshooting
Continued on next page
10-11