Limas11, Uras26: Measurement Of Calibration Cells - ABB Advance Optima magnos206 Operator's Manual

Ao2000 series gas analyzers
Table of Contents

Advertisement

Limas11, Uras26: Measurement of Calibration Cells

Definition
When should
calibration cells be
measured?
Before measuring the
calibration cells
Test Gas
Menu Path
42/24-10 EN Rev. 9
The measurement of a calibration cell in the Limas11 and Uras26 analyzer modules
means:
Determining what calibration cell "deflection" is equivalent to the test gas
calibration reading. This "deflection" is stored as the calibration cell "set value".
We recommend measuring the calibration cells once a year.
We recommend measuring the calibration cells
• after end-point calibration of a sample component with test gas or
• after any change in measurement range limits (see "Limas11, Uras26: Notes for
Changing Measurement Range Limits" section, page 7-A-3) or
• after a relinearization (see "Limas11, Uras26: Relinearization" section, page 9-8).
Prior to measuring the calibration cells, the zero and end points of the applicable
sample components must be calibrated with test gases.
During calibration cell measurement the zero gas supply must be turned on.
MENU
Maintenance/Test
Measure cal. cell
Chapter 9: Inspection and Maintenance
Analyzer spec. adjustm.
9-7

Advertisement

Table of Contents
loading

Table of Contents