Magnos206: Substitute Gas Calibration
Example
Purity Measurement
of CO
2
Calibration with
Substitute Gases
Measurement Ranges
Calibration Parameters
Calibration
Other Measurement
Tasks
8-C-10
Substitute gas calibration of the Magnos206 analyzer module is described using
the example of purity measurement of CO
For CO
purity measurements, the smallest concentrations of O
2
CO
, e.g. 0-1 Vol.-% O
2
2
Since O
in CO
is not available as a test gas and, due to the O
2
2
by CO
(see "Magnos206 Operating Specifications" section, page A-2-4), the
2
analyzer module is factory-set for calibration with substitute gas.
In this case an additional measurement range is calibrated for 0-25 Vol.-% O
N
(N
and O
/N
mixtures are available practically everywhere).
2
2
2
2
Component 1
O
in CO
2
Component 2
O
in N
2
Calibration Method
Zero Component
Span Component
• Calibrate zero with CO
• Calibrate span with dried air (containing 20.96 Vol.-% O
For other measurement tasks select the test gases and measurement ranges in a
similar manner according to the sample gas composition.
Chapter 8: Gas Analyzer Calibration
.
2
in CO
.
2
Meas. Range 1
2
Meas. Range 2
Meas. Range 3
Meas. Range 4
(Substitute Gas)
Meas. Range 1
2
Substitute Gas Calibration
Component 1
Component 2 Meas. Range 1
Component 2 Meas. Range 1
(component 1) or N
(component 2).
2
2
are measured in
2
zero shift caused
2
in
2
0-1 Vol.-%
0-15 Vol.-%
0-25 Vol.-%
0-100 Vol.-%
0-25 Vol.-%
Meas. Range 1 or
).
2