System Status: Status Messages

Where are status
messages
generated?
User-Configured
Status Messages
How are status
messages
processed?
(see Figure 10-1)
Figure 10-1
Status Message
Processing
10-2
Status messages are generated
• By the gas analyzer, i.e.
• The system controller (signal processing, calibration, system bus)
• The analyzer modules
• The gas module
• The temperature and pressure regulators
• The I/O modules and external I/O devices
• By peripherals, for example
• The system cooler
• Other sample gas conditioning modules
Status messages are automatically generated by the gas analyzer and by peri-
pherals.
Additionally, by configuring the Message generator function block, status
messages from the analyzer and peripherals can be linked into the status
message processing system (see Technical Bulletin 30/24-200 EN "AO2000
Function Blocks – Descriptions and Configuration" for details).
• Status messages are shown on the screen and recorded in the log.
• Status messages set a corresponding status signal (overall status or individual
status).
• Status signals are indicated using status LEDs and output via the system
controller digital outputs.
Component
Module
System
Peripheral
Chapter 10: Status Messages, Troubleshooting
Status Message
sets
Status Signal
Screen
Log
LEDs
Digital
Outputs
Continued on next page
42/24-10 EN Rev. 9