Manual Calibration,
Waiting Period
Following Manual
Calibration
Calibration Data
Manual Calibration of
an Analyzer Module
42/24-10 EN Rev. 9
continued
If the Output Current Response parameter is set to Hold, current output is
halted for a specific time to allow the measurement value to stabilize after manual
calibration.
This interval is:
Test gas → Sample gas purge time + 4 x T90 or
Test gas → Sample gas purge time + 1 x T90-1 + 3 x T90-2.
The waiting period is the same as that following automatic calibration (see
"Automatic Calibration" section, page 8-A-5).
Setting the calibration data is described on page 8-B-2.
Manual calibration of an analyzer module is described on page 8-D-1.
Chapter 8: Gas Analyzer Calibration