Power Supply Line Connection

What materials are
needed?
Figure 4-15
Power Supply
Connection
Power Supply Line
Connection
Potential
Compensation
Connection
4-18
The gas analyzer is supplied with a power cord and a separate two-prong
grounded plug. The power cord is 5 meters (16.4 feet) long and has a three-prong
grounded-instrument plug for connection to the power supply.
If the power cord supplied is not used, make your selection of a suitable cable
based on line length and planned current load.
The illustration shows the pin side of the plug in the power supply.
L
PE
N
Step
Action
1
Make sure the power supply leads have an adequately dimensioned
protective device (breaker).
2
Install a breaker in the power supply line or a switched receptacle
near the gas analyzer to make sure the gas analyzer can be
completely separated from the power source. Mark the breaker so
that its relationship to the protected device is clear.
3
Connect the power cord supplied to the –X01 power supply
connector in the electronics module and secure it with the clip.
4
Connect the other end of the power cord to the power source.
The gas analyzer may start when the power supply is
connected.
The electronics module and the analyzer modules have a potential compensation
connector designated by the symbol
for the installation of a matching screw or clamp.
Use this connection to link each module to the building's potential compensator in
accordance with local regulations.
Chapter 4: Electrical Connection
. The connector has M5 internal threads