Profibus Module: Electrical Connections - ABB Advance Optima magnos206 Operator's Manual

Ao2000 series gas analyzers
Table of Contents

Advertisement

Profibus Module: Electrical Connections

Figure 4-2
Profibus Module
Connection Diagram
4-2
RS485 Interface:
1 –
not used
2 M24
24 V Output Ground
3 RxD/TxD-P
Receive / Transmit Data Plus, B-Line
4 –
not used
5 DGND
Data Transmission Potential (Reference Potential for VP)
6 VP
Supply Voltage Plus (5 V)
7 P24
24 V Output Voltage Plus, max. 0.2 A
8 RxD/TxD-N
Receive / Transmit Data N, A-Line
9 –
not used
Design: 9-pin Sub-D female connector
MBP Interface (non-intrinsically safe):
1 +
2 Shield
3 –
4 not used
Design: 4-pin terminal strip for stranded or solid conductors with a maximum
2
section of 1 mm
(17 AWG). Note: Observe the information on needed materials
(see page 4-13).
Chapter 4: Electrical Connection
RS485
MBP
1
1
2
6
3
4
42/24-10 EN Rev. 9