Pump Activation, Pump Output Adjustment

Pump Activation
Pump Output
Adjustment
Menu Path
Figure 9-13
Pump Output
Adjustment
9-26
The pump installed in the internal gas module and external pumps connected to
appropriately configured digital outputs can be activated and deactivated
manually, for instance in emergency situations.
An emergency shutdown cannot be bypassed by automatic calibration.
The output of the pump installed in the internal gas module can be adjusted
manually only when the pump is in operation.
If there is a flow sensor in the gas module, its reading is shown when adjusting the
pump output.
MENU
Maintenance/Test
Chapter 9: Inspection and Maintenance
Analyzer spec. adjustm.
Pump
42/24-10 EN Rev. 9