ABB Advance Optima magnos206 Operator's Manual page 62

Ao2000 series gas analyzers
Table of Contents

Advertisement

Signal, Control and Interface Line Connection,
Wiring Connection
4-14
IP20 Version of System Housing:
Step
Action
1
Connect the wiring to the connector housing according to the
connection diagrams.
2
Connect the connector housings to the terminal strips.
Version of System Housing with connection box:
Step
Action
1
Run the cables through the connection box openings.
M20: Remove the plug from the insert; the ring remains in place as a
gasket and strain relief.
M32: Remove the plugs from the threaded connections. Press the
insert from the accessory kit over the cable; seal off the free
opening with the pin from the accessory kit.
2
Connect the wiring to the connector housings according to the
connection diagrams.
3
Connect the connector housings to the terminal strips.
Mark the cables or the connector housings of the I/O modules so that their
respective relationship to the I/O modules is clear.
Chapter 4: Electrical Connection
continued
42/24-10 EN Rev. 9