ABB Advance Optima magnos206 Operator's Manual page 160

Ao2000 series gas analyzers
Table of Contents

Advertisement

Limas11: Notes for Calibrating,
Limas11 with Internal
Cross-Sensitivity
Correction
Wait Until End of
Warm-Up Phase
8-C-6
During calibration cross-sensitivity and / or carrier gas corrections by other meas-
urement components are switched off. Therefore, the following notes should be
observed:
For zero-point calibration all sample components must be calibrated in the
following sequence:
• First that sample component which is not corrected,
• Afterwards that sample component which is affected by the smallest number of
corrections,
• Up to that sample component which is affected by the largest number of
corrections.
Example:
Sample components
Cross-sensitivity correction
Sequence for zero-point calibration
For end-point calibration likewise all sample components must be calibrated. In
doing so, a corrected sample component should be calibrated only using a test
gas containing no components which cause cross-sensitivity, i.e. consisting only
of the measurement component and an inert gas like N
The Limas11 analyzer module should only be calibrated after the warm-up phase,
i.e. approx. 1.5 hours after connecting the power supply.
Chapter 8: Gas Analyzer Calibration
continued
NO, SO
, NO
2
2
NO by SO
and NO
,
2
2
SO
by NO
,
2
2
NO
not corrected.
2
st
nd
rd
1
NO
– 2
SO
– 3
NO.
2
2
.
2
42/24-10 EN Rev. 9