Adding a System Module
Adding a New
Analyzer Module or a
New I/O Device
Adding a New
I/O Module
42/24-10 EN Rev. 9
• If no system module is configured or an added system module is not yet config-
ured in a gas analyzer, the NEW softkey will appear on the screen. Press this
softkey to go directly to the Setup System Modules menu.
• The procedure for adding an analyzer module and an external I/O device differs
from the procedure for adding an I/O module (see the following instructions).
Step
Action
Select the Set Up System Modules menu item.
1
The list of the system modules existing in the system is displayed.
Press the NEW softkey.
2
3
Enter the 14-digit serial number for the new system module.
In the list the new system module will be shown with a New status.
4
Save the configuration change by pressing ENTER or discard it by
5
pressing Back.
Step
Action
Select the Set Up System Modules menu item.
1
The list of the system modules existing in the system is displayed.
2
Select the I/O module which has been added and automatically
recognized by the gas analyzer and press the NEW softkey.
When adding a digital I/O module: Press the FB APPL. softkey and
3
select a function block application.
In the list the new system module will be shown with a New status.
4
Save the configuration change by pressing ENTER or discard it by
5
pressing Back.
When retrofitting a Profibus module, it must always be installed in the lowest slot
X20 / X21.
Chapter 7: Gas Analyzer Configuration