Digital I/O Module: Electrical Connections

Figure 4-7
Digital I/O Module
Connection Diagram
42/24-10 EN Rev. 9
24V
24V
24V
24V
DI1-DI4
Digital Inputs:
Optocoupler with internal 24 VDC power supply. Control with floating
contacts, with external voltage 12–24 VDC or with open collector
drivers PNP or NPN..
DO1-DO4
Digital Outputs:
Floating double-throw contacts, max. contact load rating 30 V/1 A
Relays must at all times be operated within the specified data range.
Inductive or capacitive loads are to be connected with suitable
protective measures (self-induction recuperation diodes for inductive
loads and series resistors for capacitive loads).
Design: 2x12-pin terminal strip for stranded or solid cable with a maximum section
2
of 1 mm
(17 AWG). Observe the information on needed materials (see page 4-13).
"Standard Terminal Connections" section (see page 4-9) contains information on
standard terminal assignment. Please observe also the specifications in the
analyzer data sheet. Relays are shown in the unpowered state. The unpowered
state is the failure mode ("fail safe").
Chapter 4: Electrical Connection
DI4
DI3 DI2 DI1
DO4
DO3 DO2 DO1
1
3
5
7
9 11 13 15 17 19 21 23
2
4
6
8 10 12 14 16 18 20 22 24
1
DI4 -
2
+
12-24 V
3
GND
5
DI3 -
6
+
12-24 V
4
GND
7
DI2 -
8
+
12-24 V
9
GND
11
DI1 -
12
+
12-24 V
10
GND
13
DO4 NO
max. 30 V / 1 A
15
Common
17
NC
14
DO3 NO
max. 30 V / 1 A
16
Common
18
NC
19
DO2 NO
max. 30 V / 1 A
21
Common
23
NC
20
DO1 NO
max. 30 V / 1 A
22
Common
24
NC
Continued on next page
4-7