Configuring Profibus
Menu Path
Figure 7-C-4
Profibus
Configuration
Parameter
Profibus address
Profibus type
Profibus baudrate
Profibus map
Profibus restart
7-C-14
MENU → Configure → System → Network → Profibus
Selection
1 to 126
Profibus DP
Connection to the RS485 interface
Profibus PA
Connection to the MBP interface (non-intrinsically safe)
RS485 interface
automatically,
9600 Baud, 19200 Baud, 93750 Baud, 187.5 KBaud,
500 KBaud, 1500 KBaud, 3000 KBaud, 6000 KBaud
MBP interface
set to 31250 Baud
Profibus inputs
Measured values, Bus analog outputs, Analog inputs, Analog
outputs, Digital inputs, Bus digital outputs, Digital outputs
Profibus outputs
Bus analog inputs, Bus digital inputs
Warm start
With Warm start the Profibus stack is reset, comparable with a
power off/on.
Cold start
With Cold start, all the parameters which are stored in the
Profibus stack as store parameters are reset to the default
value.
For additional Profibus information see the "AO2000 Profibus DP/PA Interface"
technical bulletin (Publication No. 30/24-315 EN).