Display Features,
Function Blocks as
Sources
Display Element
Positioning on the
Screen
Figure 7-D-1
Position Numbering
Views
7-D-2
continued
The values of all the function blocks in the system can be configured as a source
for the display. The source of the display of value entries or key entries is also a
function block that was created when configuring the display elements. The
display of the function block value is independent of the other links of the function
block.
Note: All measuring components, auxiliary quantities, current outputs and current
inputs exist as function blocks in the system, i.e., all of these measured quantities
are displays of function blocks in the system.
The display elements can be represented in two sizes. A maximum of three large
and six small display elements can be represented on a page. Large and small
display elements can be mixed with each other. The positions are numbered as
shown in Figure 7-D-1. The numbering of the positions corresponds to the arrange-
ment of the number keys next to the display.
8
5
2
The following views are available for screen configuration:
• Display overview (see page 7-D-3)
• Page overview (see page 7-D-4)
• Parameter overview (see page 7-D-5)
Chapter 7: Gas Analyzer Configuration
7
9
4
6
1
3