Air Pressure Correction

Air Pressure Effect
Measures to Minimize
Air Pressure Effect
Which analyzer
modules have a
pressure sensor?
Air Pressure Values
9-28
A specific amount of change in air pressure will result in a specific change in a
measurement value, depending on the measurement principle employed by the
analyzer module.
Air pressure effect can be minimized by:
• Installing a pressure sensor in the analyzer module (this can only be done at the
factory) or
• Entering the current atmospheric pressure as a correction value.
Analyzer Module
Uras26, Limas11, Caldos27
Caldos25
Magnos206, Magnos27
MultiFID14
Diagnostic/Information
Use the MENU
item and select the appropriate analyzer module to determine if a pressure sensor
is installed.
Operating Altitude
meters above
mean sea level
–200
–100
±0
+100
200
300
400
500
600
700
800
900
1000
1100
1200
1300
1400
1500
1600
1700
1800
1900
2000
Chapter 9: Inspection and Maintenance
Mean Air Pressure
hPa (mbar)
1037
15.04
1025
14.87
1013
14.69
1001
14.52
989
14.34
977
14.17
965
14.00
955
13.85
943
13.68
932
13.52
921
13.36
909
13.18
899
13.04
888
12.88
877
12.72
867
12.57
856
12.42
845
12.26
835
12.11
825
11.97
815
11.82
804
11.66
793
11.50
Pressure Sensor
built-in at the factory
not necessary
built-in at the factory as an option
cannot be built in
System overview menu
psi mm Hg (Torr)
778
769
760
751
742
733
724
716
707
699
691
682
674
666
658
650
642
634
626
619
611
603
595
42/24-10 EN Rev. 9
in Hg
30.63
30.28
29.92
29.57
29.21
28.86
28.50
28.19
27.84
27.52
27.21
26.85
26.54
26.22
25.91
25.59
25.28
24.96
24.65
24.37
24.06
23.74
23.43