Changing the Number of Decimal Places
Menu Path
Selection
Procedure
Number of Decimal
Places
7-A-4
MENU → Configure → Component specific → Measurement range
(→ Select component) → ...
All measurement ranges configured (at the factory) for a sample component are
displayed.
Select the measurement range with the arrow keys, press SET DECIMAL, set the
number of decimal places with the softkeys and confirm the selection by pressing
ENTER.
The setting only affects the values indicated on the screen.
The number of decimal places cannot be increased.
The altered number of decimal places is shown on the screen after switching to
measurement mode.
When the screen displays the measured value in physical units (e.g. ppm) the
number of places after the decimal point depends on the size of the set measur-
ing span:
Span (in physical units)
Places after the decimal point
For the display of the measured value as a percentage of the whole measuring
range (%Span) two places are always shown after the decimal point.
The number of decimal places when setting the parameters is the same as in the
display in measuring mode.
Chapter 7: Gas Analyzer Configuration
≤ 5 > 5 to ≤ 50
> 50 to ≤ 500 > 500
3
2
42/24-10 EN Rev. 9
1
0