Uras26 Problems - ABB Advance Optima magnos206 Operator's Manual

Ao2000 series gas analyzers
Table of Contents

Advertisement

Uras26 Problems

Temperature Problem
Faulty temperature
sensor or heater
connections
Defective thermal link • Check thermal link continuity and replace if necessary.
Unstable Readings
Vibration
Gas path leakage
Loss of sensitivity
Uneven emitter
modulation
10-22
• Check the connecting lines and plugs.
• Check the line seating in the insulated jackets.
• Take measures to reduce vibration.
Permissible vibration levels: for analyzer max. ±0.04 mm at 5 to 55 Hz, 0.5 g at
55 to 150 Hz; when installed in cabinet max. 0.01 ms
• Increase the low pass time constant T90 (see page 7-A-5 for instructions).
• Check the integrity of the analyzer module gas paths and (if applicable) of the
lines to the gas module (see page 9-2 for instructions).
• Check the sensitivity variation:
Indication < 75%: The "Maintenance request" status signal appears.
The detector involved will need to be changed soon.
Indication < 50 %: The "Failure" status signal appears.
Replace the detector involved.
• Remove the emitter.
The emitter temperature is approx. 60 °C in the thermostat version of the
Uras26!
• Check if the chopper wheel turns smoothly.
• Check the clamp ring seating.
• The chopper wheel should not extend beyond the notch.
• Have the emitter and modulator assembly checked by the service department.
Chapter 10: Status Messages, Troubleshooting
–2
at 0.1 to 200 Hz.
42/24-10 EN Rev. 9