ABB Advance Optima magnos206 Operator's Manual page 18

Ao2000 series gas analyzers
Table of Contents

Advertisement

Installation Location Requirements,
Installation Location
Altitude
Vibrations
1-2
The maximum installation location altitude is 2000 m.
If the gas analyzer is installed in a cabinet the maximum acceleration amplitude is
–2
0.01 ms
in a frequency range of 0.1 to 200 Hz.
If the gas analyzer is not installed in a cabinet the following data for the individual
analyzer modules apply.
Analyzer Module
Caldos25
Caldos27
Limas11
Magnos206
Magnos27
Uras26
Note: Compliance with the metrological data can only be assured if data on
vibration amplitude and frequency range at the installation site are available and
suitable means are employed to decouple the analyzer.
Chapter 1: Preparing the Installation
continued
Vibrations
max. ±0.04 mm at 5 to 30 Hz
max. ±0.04 mm at 5 to 55 Hz, 0.5 g at 55 to 150 Hz
max. ±0.04 mm at 5 to 55 Hz, 0.5 g at 55 to 150 Hz
max. ±0.04 mm at 5 to 20 Hz
max. ±0.04 mm at 5 to 60 Hz
max. ±0.04 mm at 5 to 55 Hz, 0.5 g at 55 to 150 Hz,
slight transient effect on sample value in the region
of the beam modulation frequency
42/24-10 EN Rev. 9