Warm-Up Phase

Warm-Up Phase
Warm-Up Phase
Duration
5-4
The duration of the warm-up phase depends on which analyzer module is
installed in the gas analyzer.
Analyzer Module
Warm-Up Phase Duration
Caldos25
1.5 hours
Caldos27
Approx. 30 / 60 minutes for class 1/ 2 measurement ranges
Limas11
Approx. 2.5 hours
≤ 1 hour
Magnos206
Magnos27
2 to 4 hours
Uras26
Approx 0.5 / 2 hours without / with thermostat
1) See "AO2000 Series" Data Sheet (publication number 10/24-1.20 EN) for
class details.
• The warm-up phase can take longer if the gas analyzer was not brought to room
temperature before the power supply was activated.
• During the warm-up phase measurement values can be outside the ranges
specified in the data sheet.
The warm-up phase is over when the measured value drift indication is accept-
able. This depends on the size of the measurement range.
Chapter 5: Gas Analyzer Start-Up