Calibration Methods,
Overview
Quantity
Calibration
SC
MR Method
1
1
Test Gas /
Single
≥ 1
> 1
Test Gas /
Common
≥ 1
> 1
Substitute
Gas
Setting the
Calibration Method
(see Figure 8-A-7)
Figure 8-A-7
Setting the
Calibration Method
42/24-10 EN Rev. 9
continued
The following table summarizes the various calibration methods.
To configure ...
Measurement ranges for
zero and span calibration
Components and
measurement ranges for
zero and span calibration
SC = Sample and Substitute Gas Components
MR = Measurement Ranges per Component
The calibration method can be set separately for manual, automatic and externally
controlled calibration.
For common and substitute gas calibration the sample ranges for start- and end-
point calibration of all three types of calibration control are adjusted jointly.
For substitute gas calibration the zero and span calibration components should
also be set.
Test gas
Component
Common
calibration
Zero range
Chapter 8: Gas Analyzer Calibration
Calibrate ...
• Zero
• Span
individually for each sample
component and range
• Zero in one measurement
range
• Span in another range
for each sample component
• Zero in one component
measurement range
• Span in one range for
another component
for each detector
Calibration
method
Single
calibration
Span range
Calibration affects ...
Measurement range
All sample component
measurement ranges
All detector components
and measurement ranges
Substitute
gas
Zero
Span
component
component
Zero range
Span range
8-A-11