ABB Advance Optima magnos206 Operator's Manual page 196

Ao2000 series gas analyzers
Table of Contents

Advertisement

Limas11: Amplification Optimization,
Menu Path
Procedure
9-24
MENU
Maintenance/Test
Amplification optimization
Step
Action
1
Turn on the zero gas supply.
If a solenoid valve is used to switch to zero gas, the supply will be
activated automatically.
Select the Amplification optimization menu item.
2
3
Select the first sample component for which status message No. 301
is present.
Press the New Lamp or Optimize or Optimize All softkey.
4
After pressing New Lamp a window will be appear for entering the
serial number of the new lamp. After this the amplification optimiza-
tion procedure will run automatically for all sample components and
cannot be interrupted.
5
Wait for measurement value readings to stabilize and end the amplifi-
cation optimization procedure by pressing ENTER.
Confirm the amplification optimization by pressing ENTER (the zero
6
point will be automatically aligned) or reject the result by pressing
Back or Meas.
7
Repeat steps 3-6 for all sample components for which status message
No. 301 is present..
Chapter 9: Inspection and Maintenance
continued
Analyzer spec. adjustm.
42/24-10 EN Rev. 9