Power Supply Line Connection to the Analyzer Module

24-VDC Connecting
Cable
Figure 4-12
24-VDC Connecting
Cable
Figure 4-13
24-VDC Connection
Figure 4-14
Connecting 24-VDC
Power from the
Central Unit Power
Supply to a Separate
Analyzer Module
4-16
• The following information and instructions should be followed when connecting
the 24-VDC power supply to an analyzer module that is not installed in the
central unit but in a separate system housing.
• Information in the sections "Power Supply Information", page 1-7, and "Power
Supply", page 1-8, should also be followed.
If an analyzer module is not installed in the central unit but in a separate system
housing a 5-meter (16.4-feet) long cable (2 x 0.5 mm
supplied.
The receptacle on one cable end is designed to connect to the 24-VDC male plug
on the analyzer module or the I/O base module (see Figure 4-13).
The wires on the free end of the connecting cable are intended for connection to
• The power supply filter -Z01 in the central unit (for an example see Figure 4-14) or
• An external power supply
The illustration shows the pin side of the analyzer module
3
plug and thus the solder side of the matching female jack.
4 2
1
CU
EM
-Z01
115 / 230 V AC
AM
Analyzer Module
CU
Central Unit
Chapter 4: Electrical Connection
24 V DC
2
(0.5 mm ,
<
5 m)
BUS
EM
Electronics Module
PS
Power Supply
2
, see Figure 4-12) will be
BN +
BU -
-Z01 Power Supply Filter
BUS System Bus
Continued on next page