Changing Analog Output Current Range

Method
Menu Path
Figure 9-14
Analog Output
Function Block
Parameters
Changing Current
Range
Selection
42/24-10 EN Rev. 9
The current range of the individual analog outputs can be modified by initializing
the applicable Analog output function blocks. Technical Bulletin "AO2000 Func-
tion Blocks – Descriptions and Configuration" (Publication No. 30/24-200 EN)
contains a detailed description of the function block.
MENU
Configure
Analog output
The current range is changed using the User range parameter.
The current range choices are 0-20 mA, 2-20 mA and 4-20 mA.
The output signal cannot be lower than 0 mA.
Chapter 9: Inspection and Maintenance
Function blocks
Outputs
9-27