Gas Analyzer Installation

What materials are
needed?
Installation
2-6
CAUTION!
A system housing with an electronics module and an analyzer module
weighs from 18 to 23 kg (40 to 50 pounds). The following points should be
observed:
• Two persons are needed for installation.
• The location (e.g. bay, 19-inch rack, wall) must be capable of supporting the
gas analyzer's weight.
• The 19-inch housing must be supported with rails in the bay or rack.
• Neither the 19-inch nor the wall-mount housings use hinges to secure the
housing cover. The cover can drop when opened.
Quantity
Fastener (not supplied)
19-inch housing:
4
Oval-head screws (Recommendation: M6; this depends on the
cabinet / shelf system)
1 pair
Rails (Design depends on cabinet / shelf system)
Wall-mount housing:
4
M8 or M10 bolts
Install the system housing in the cabinet / shelf or on the wall with the required
fasteners. Observe the dimensional diagrams and the additional notes on page 2-4.
Chapter 2: Gas Analyzer Unpacking and Installation
42/24-10 EN Rev. 9