Removing A System Module 7 - ABB Advance Optima magnos206 Operator's Manual

Ao2000 series gas analyzers
Table of Contents

Advertisement

Removing a System Module
Sequence when
Removing System
Modules
Removing an Installed
System Module
without a Replacement
42/24-10 EN Rev. 9
When removing system modules from the gas analyzer proceed according to the
following sequence:
1. Remove the system module in the software (see below for instructions)
2. Dismount the system module from the gas analyzer.
Step
Action
Select the Set Up System Modules menu item.
1
The list of the system modules existing in the system is displayed.
2
Select the system module to be removed (and not replaced).
Press the DELETE softkey.
3
This will irretrievably remove the system module's parameter
settings and function block configuration!
In the list the deleted system module's status will change to Deleted.
4
Save the configuration change by pressing ENTER or discard it by
5
pressing Back.
Chapter 7: Gas Analyzer Configuration
7-C-9

Advertisement

Table of Contents
loading