ABB Advance Optima magnos206 Operator's Manual page 186

Ao2000 series gas analyzers
Table of Contents

Advertisement

Limas11: Quartz Sample Cell Cleaning,
Figure 9-7
Quartz Sample Cells
in the Limas 11
Analyzer Module
Figure 9-8
Elbow / Pivot Fitting
Components
9-14
Step
Action
Restarting the analyzer module:
19
Close the system housing.
Light penetration during operation leads to erroneous measure-
ment values and measurement range overflows ("Intensity"
status message).
20
Turn on the gas analyzer power supply.
21
Wait for the warm-up phase to end. Start the sample gas supply.
22
Check linearity.
1
2
1
Threaded Fittings on Housing Back Wall
2
Sample Gas Tubes
3
Allen Screws (3 mm)
2
13
12
Chapter 9: Inspection and Maintenance
Continued
3
4
5
11
10
9
8
7
6
3
4
4
Mounting Bracket
5
Elbow / Pivot Fittings (see
Figure 9-8 for Components)
6
Quartz Sample Cell
2
Sample Gas Tube
7
FFKM75 O ring
8
Internal Part
9
Elbow
10 Nut
11 Sealing Ring
12 Cutting Ring
13 Nut
42/24-10 EN Rev. 9