Status Messages,
No. Text
System Cooler
1100 Cooler temperature too high.
1101 Cooler temperature too low.
1102 Condensate penetration in
cooler.
1103 Flow rate inadequate in cooler.
1104 Cooler condensate level too
high.
1105 Cooler condensate level too
high.
1106 Cooler reagent level too low.
42/24-10 EN Rev. 9
continued
O E M F Reaction / Remark
x x
Sample gas feed module pump is automatically turned off.
Check the system cooler and sample gas preparation.
x x
x x
x
Check the system cooler and sample gas preparation.
x
x
Empty the condensate bottle.
x
Fill the reagent container.
Chapter 10: Status Messages, Troubleshooting