Caldos27 Operating Specifications

Linearity deviation
Repeatability
Zero drift
Sensitivity drift
Output fluctuation (2 σ) ≤ 0.5 % of smallest measurement range span at electronic T90 time = 0 sec
Detection limit (4 σ)
Flow influence
Associated gas
influence
Temperature
influence
Air pressure influence ≤ 0.25 % of span per 10 hPa for the smallest possible ranges given; in larger
Power supply
influence
Position influence
T
time
90
A-2-2
≤ 2 % of span
≤ 1 % of span
≤ 2 % of the smallest possible measurement range per week
≤ 0.5 % of the smallest possible measurement range per week
≤ 1 % of smallest measurement range span at electronic T90 time = 0 sec
The stability data apply only to measurement ranges ≥ Class 2.
≤ 0.5 % of span at a flow change of ±10 l/h. At the same sample and test gas flow
rate the flow influence is automatically compensated.
The knowledge of the sample gas composition is necessary for the analyzer
configuration.
At each point in the measurement range:
≤ 0.5 % of span per 10 °C, based on temperature during calibration;
Ambient temperature in permissible range, thermostat temperature = 60 °C
measurement ranges the influence effect is correspondingly smaller.
Working range of installed pressure sensor: p
Option: Operating altitude over 2000 m
24 VDC ± 5 %: ≤ 0.2 % of span
< 1 % of span up to 30° deviation from horizontal orientation
≤ 2 sec. with direct sample chamber connection and sample gas flow = 60 l/h
T
90
(applies to a gas analyzer with one analyzer module)
Appendix 2: Analyzer Module Operating Specifications
= 600 to 1250 hPa
abs
42/24-10 EN Rev. 9