Gas Module Problems

Flow Problem
Condensation in the
flow meter
Inadequate gas
supply
42/24-10 EN Rev. 9
• Disconnect the gas analyzer from the gas preparation system.
• Dry the flow meter by heating it and blowing it with compressed air.
• Check the operation of the upstream sample gas cooler.
• Connect the flow meter, ball chamber or pressure gauge directly to the gas
supply pump and measure the pressure or vacuum.
• Check the pump and change the membrane as necessary.
• Check and, if necessary, replace the disposable filter (see page 9-24 for
instructions).
• Check and, if necessary, replace the solenoid(s).
Chapter 10: Status Messages, Troubleshooting
10-23