To Attach A Capillary Column Adapter On An Adaptable Fid - Agilent Technologies Agilent 7890 Series Maintenance Manual

Gas chromatograph
Table of Contents

Advertisement

To Attach a Capillary Column Adapter on an Adaptable FID

WA R N I N G
C A U T I O N
Maintaining Your GC
1
Gather the following materials:
Adapter (See
"Consumables and Parts for the FID"
page 196.)
1/4- inch brass nut
1/4- inch Vespel/graphite ferrule
Column cutter
1/4- inch wrench
9/16- inch open- end wrench
Lint- free gloves
Load the
GC maintenance method
2
become ready.
Be careful! The oven, inlet, and/or detector may be hot enough to
cause burns. If the oven, inlet, or detector is hot, wear
heat-resistant gloves to protect your hands.
Wear clean, lint-free gloves to prevent contamination of parts with
dirt and skin oils.
3
Assemble the nut and ferrule onto the adapter.
1/4-inch ferrule
1/4-inch nut
Adapter
4
Insert the adapter straight into the detector base as far as
possible.
Hold the adapter in this position and finger- tighten the
5
nut.
Maintaining the FID
and wait for the GC to
11
on
203