Switching A Relay Channel - Agilent Technologies 3488 User Manual

Switch/control system
Table of Contents

Advertisement

Switching a Relay Channel

Switch modules can be used to route signals to and from your test
system. This is achieved by closing or opening the relay channels on
these modules.
From the front-panel, you can open or close one relay channel at a
time. However, over the GPIB interface, multiple relay channels can
be operated by a single command if a channel list is specified. In
addition, these open or closed states can be stored, and a stored
channel setup can be included in a scan list (see "State Storage" on
page 94).
Whenever a switch module is reset, all of the closed relay channels on
the module will be opened.
Whenever the instrument is turned on or reset, all of the closed relay
channels in the instrument will be opened.
Front-Panel Operation: Select a channel, press Open or Close.
Select a slot, press and hold Card Reset to open all channels on the
selected module.
Press Shift, then press and hold Card Reset to open all channels in
the instrument.
Remote Interface Operation: The following commands open and close
multiple channels on the modules in slots 1 and 2.
OUTPUT 709; "OPEN 101,102,207"
OUTPUT 709; "CRESET 1"
OUTPUT 709; "CLOSE 101,102,207"
Chapter 4 Features and Functions