Figure 15-1
Option 002 Frequency
Cal:
Power Meter to DUT Connection
2. Begin the self calibration of the YTF, by pressing the following softkeys:
Service
YTF Cal
YTF1 Power
3. The process takes about 15 minutes. Once complete, the unit will have
accurately determined the detector calibration to set output power.
1. Begin the self calibration of YTF2, by pressing the following softkeys:
Service
YTF Cal
YTF2 Freq
2. The process takes approximately 10 minutes to complete. Once
complete, the unit will have accurately determined the current settings of
the YTF to select any frequency between 2.4 GHz and 26.5 GHz.
HP 70427A/HP 70428A User's Guide 15-3
Adjustments