NOTE
Frequency Cal:
Power Cal:
NOTE
NOTE
Adjustments
mode. Then configure the DUT for the RF Source by using the following
softkey sequence:
Service
Misc
RF Source
or
HP 8662A
HP 8663A
If you selected HP 8662A or HP 8663A, continue by selecting the HP-IB
address with the key sequence below. If you selected Other, the RF Source is
not controlled over HP-IB, so you may skip the remaining soft key sequence
listed below. However, you must manually setup the RF Source. Set
frequency to 600 MHz, power to +5.0 dBm, and turn off all modulation.
Verify that the RF Source is within 1 dB of the desire level by using a power
meter.
HP-IB ADDRESS
NN (address of the RF Source. Range= 1 to 30)
ENTER
1. Begin the self calibration of the YTF, by pressing the following softkeys:
Service
YTF Cal
YTF1 Freq
2. The process takes approximately 10 minutes to complete. Once
complete, the unit will have accurately determined the tuning of the YTF
to select any comb line between 2.4 GHz and 26.4 GHz.
1. Connect the equipment as shown:
a: The UUT is connected to the Power Meter via the MSIB for the HP
70100A or via HP-IB for the HP 437A/B.
b: The sensor is connected to the μW LO Output with only the f-f adapter,
HP part number 5061-5311 or equivalent.
15-2 HP 70427A/HP 70428A User's Guide