Voltage tuning range of VCO: To be entered later.
VCO tune-port input resistance: 1E6
Phase detector: Internal Phase Detector, 1.2 GHz to
18 GHz.
3. Specify a calibration process
Method for determining phase detector constant: Measure
the detector constant.
Method for determining the VCO tuning constant:
Measure the VCO tuning constant.
PLL suppression: Will be verified.
4. Specify the control of signal sources
DUT: User's source, manual control.
DN CONV: Not in use.
REF SOURCE: User's source, manual control.
TIMEBASE: Not in use.
TUNE VOLTAGE: Ref source.
5. Define the graph of results
Title: To be entered later.
Minimum X coordinate: 1 Hz
Maximum X coordinate: 40 MHz
Minimum Y coordinate: –170
Maximum Y coordinate: 0
Horizontal portion of display to be used: 100%
Vertical portion of display to be used: 100%
Graph type: Single sideband phase noise (dBc/Hz)
6. Enter the spec lines
To be entered later
Step 2 - Reference source initial configuration
1. Make the reference source the active MMS module (ACT indicator lit):
a. Press the
DISPLAY
b. Press the
NEXT INSTR
indicator lights.
2. Press the
INSTR PRESET
3. Access the Source screen: if the module is an HP 70427A or HP 70428A
Option 002 press the
Test 2 - Phase Noise and Spurs (mW LO)
key.
softkey until the reference source's ACT
key.
softkey.
Source
HP 70427A/HP 70428A User's Guide 14-9