Performance Tests
Test 3 - AM Noise and Spurs (mW LO)
2. Measure and record the DUT power
Figure 14-4
DUT Power Measure and Record
Make sure the AM detector is connected to the DUT's output, and that the
HP 3561A is connected to the detector's output via a Tee connector.
Record the DC voltage measured across the AM detector's output with the
HP 3561A as described in 1 above.
3. Set up the calibration source
14-28 HP 70427A/HP 70428A User's Guide
Frequency: 1.2 GHz
Power: +10 dBm
Modulation: OFF